<ErrorNumber>15080</ErrorNumber>
<ErrorFamily>Autoloader</ErrorFamily>
<ErrorMessage>
ALE: timeout moving trailing edge cylinder
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>All</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
Check for faulty TEC sensors by using the "tec" command from the diagnostic. If the sensor is
faulty, use "set al te 0" or "set al tes 0" to bypass the sensor and get the customer going.
Arrange for field service.  If the sensor is OK, test the LEC actuator by using "tec on" or
"tec off" and see if any motion occurs.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Clear obstruction near TEC actuator and select RESUME.
</text></DescriptionText>
</OperatorAction>
<Cause>
Faulty TEC sensor
</Cause>
<Cause>
Faulty TEC actuator
</Cause>
<SeeAlso>
<UnnumberedList>
NVS: al te cal
<Link><LinkAddress>../nvs/al_cal_tes.xml</LinkAddress><LinkTitle>NVS: al tes</LinkTitle></Link>
<Link><LinkAddress>../commands/tec.xml</LinkAddress><LinkTitle>Command: tec</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:32:51 2008 PST
</Updated>
<Links>
</Links>
</Error>